Page 106

2. Select the Displacement channel.
3. Click the Limits tab.
4. Set the Ch1 Displacement Upper Action and Lower Action to Disabled.
Switch to Running Max/Min Displacement and Force Meters
1. Switch to a Running Max/Min Displacement meter.
Because you are no longer imparting a waveform on the specimen, change the displacement Peak/Valley
meters to Running Max/Min Meters to monitor history during specimen removal.
a) In the Station Manager application main window, click the Meters button.
b) Click in the title bar of each Peak/Valley meter and change Meter Type to Running Max/Min.
2. Switch to a Running Max/Min Force meter.
106 | MTS Landmark Test System Operation
Running the Example Tension Test