and the crosshead. This typically includes the lower grip, the specimen, the upper grip, and the force
sensor (load cell), as shown.
Typical Components in the Load Train
Description
Item
Force Sensor (Load Cell)
1
Spiral Washers
2
Upper Grip
3
Specimen
4
Lower Grip
5
Connector Stud
6
Actuator
7
Positioning the Crosshead to Install the
Specimen
It is best practice to position the crosshead to accommodate the specimen and fixturing while ensuring
minimal movement away from the mechanical center. In general, once you position the crosshead, additional
adjustments are not needed unless a component in the load train changes dimensions substantially. On
60 | MTS Landmark Test System Operation