The current line water temperature can be queried manually if the water supply is

open through the

&RROLQJ ZDWHU

menu item (see Section 11.1 “Maintenance menu”)

If the line water temperature is subject to large fluctuations, we recommend that you

work with a cooler unit.

h

&RQQHFWLQJ WKH PDLQV SRZHU OLQH

Check the voltage information on the rating plate of the calorimeter with the data for

your power supply network.

Connect the mains line with the calorimeter or with the power source.

&RQQHFWLRQ R[\JHQ VXSSO\

Installation

The 2 m long O

2

hose supplied with the system is designed for a maximum pressure

of 40 bar at room temperature. This can either be fed directly to the oxygen cylinder

or to another oxygen connection point with a pressure reducer (smallest permitted

bending radius 80 mm).

The C 29 reducer valve (accessory) for the oxy-

gen cylinder features an R ¼ inch thread for con-

necting the pressure hose.

The pressure hose is supplied with an adapter for

American pressure reducers.

i

Make sure that the oxygen supply line is connected to the C 2000 calorimeter as

described under Point

e

. Then connect the O

2

line to the laboratory oxygen supply

end.

7KH SUHVVXUH RI WKH R[\JHQ VKRXOG EH EDU EXW PXVW QRW LQ DQ\ FDVH H[FHHG

EDU

Keep all gas lines and screw connections that carry oxygen free from grease.

Observe the accident prevention requirements applicable to the activity and the

work station. Wear personal safety equipment.

Close the main valve on the oxygen supply when work is complete.

Only carry out maintenance work when the system is depressurised.