6hwxs dqg frpplvvlrqlqj, 6hwxs – IKA C 2000 control User Manual
Page 25
IKA
-WERKE C 2000 basic / control
Ver. 04 08.07
3DJH
6HWXS DQG FRPPLVVLRQLQJ
You have unpacked the components of the C 2000 calorimeter system and they are
located where they will be set up (see Chapter 3, Section 3.2, “Set-up location”).
Now perform the following steps:
6HWXS
All connections for the supply lines and peripheral devices are located on the rear
side of the device.
c
&RQQHFWLQJ WKH R[\JHQ VXSSO\ OLQH
Screw the pressure hose (O
2
line) with the M8x1 cap screw onto the oxygen con-
nection adapter of the measuring cell (fork wrench SW 10, included with delivery)
and connect the hose to your oxygen supply on the laboratory side. On the pressure
side, the hose is designed with an R ¼” screw thread. A ¼” NPT adapter piece is
delivered with all hoses. In some circumstances this must be used to replace the ¼”
screwing piece (users in the USA).
Please refer to chapter 6.2 for details how to connect to oxygen supplies.
1
Connecting cable connection
2
Sample rack connection
3
Keyboard connection
4
Scale connection
5
PC connection
6
Fuses
7
Monitor connection
8
Printer connection
9
Mains power connection
10
Water drain adapter
11
Oxygen connection adapter
12
Water outlet adapter
13
Water intake adapter