Page 79

background image

Chapter 7

______________________________________________________________________

_____________________________________________________________________________

BE6025-15

1412i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 79 of 202

Cross Interference: if any “interferent” other than water vapour is

present in the ambient air-sample, the Monitor can cross compensate

for the interference caused by its presence provided that a selective

optical filter is installed in the monitor to measure the concentration

of interferent gas present.

Gas Monitor Display / Averaging Period (1412i stand-alone use):

this enables you to decide if instantaneous values or averaged values

are displayed on the gas monitors screen.

If average values are desired, then the example below illustrates

how the values are calculated and displayed.

Let us suppose that the Monitor has been taking measurements

every 2.5 min., over a period of 30 min., of gas A (acetone), and

the following results (unshaded area) were obtained and recorded

in the display memory, for stand-alone users.

A: ACETONE

Averaged Values

0

11:42:30

1

66.72 mg/m

3

66.72

mg/m

3

0

11:45:00

2

154.90 mg/m

3

110.9

mg/m

3

0

11:47:30

3

271.30 mg/m

3

164.3

mg/m

3

0

11:50:00

4

415.90 mg/m

3

227.2

mg/m

3

0

11:52:30

5

588.70 mg/m

3

299.5

mg/m

3

0

11:55:00

6

789.70 mg/m

3

444.1

mg/m

3

0

11:57:30

7

1.019

g/m

3

616.9

mg/m

3

0

12:00:00

8

1.276

g/m

3

817.9

mg/m

3

0

12:02:30

9

1.562

g/m

3

1.047

g/m

3

0

12:05:00

10

1.876

g/m

3

1.304

g/m

3

0

12:07:30

11

2.218

g/m

3

1.159

g/m

3

0

12:10:00

12

2.589

g/m

3

1.904

g/m

3

Let us suppose that the averaging period was set to 10 min. When

the above measurement results are displayed on the screen and

the averaging function is activated then each measurement will be

averaged over the previous 10 min. and the averaged

measurement results will replace the individual measurement

results (averaged results are shown in the shaded area in the list

above). For example: measurement number 7 would be 616.9

mg/m

3

. This is the average of the 5 measurements taken in the

10 min. period ending at 11:57:30. Measurement number 12

would be 1.904 g/m

3

which is the average of the 5 measurements

taken in the 10 min. period ending at 12:10:00.