Chapter 13

______________________________________________________________________

_____________________________________________________________________________

BE6025-15

1412i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 186 of 202

6. Insert the Optical Filters and fit the locking spring. Check that the

locking spring is fitted correctly around the groove of the optical

filter.

7. Reinsert the Filter Wheel with the mark “Face IR-Source” towards

the IR-source.

8. Note the UA number of each optical filter and its position in the

filter carousel (for example, UA 0983 in position “B”, UA 0988 in

position “D” and so on).

9. Enter the correct UA number of the optical filter installed in each

position (“A” to “E”) of the filter carousel –

section 12.3.3

provides

details. This is a very important step as it determines the set of

Optical Filter Factors, which will be used during the calibration of

each optical filter. If the UA number is not correctly entered then

the filter cannot be calibrated to measure any gas accurately.