beautypg.com

Gas concentrations – single-point calibration, Gas concentrations – two-point calibration – LumaSense Technologies INNOVA 1412i User Manual

Page 165

background image

Chapter 12

______________________________________________________________________

_____________________________________________________________________________

BE6025-15

1412i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 165 of 202

The question about whether to perform a single- or a two-point span

calibration is difficult to answer, as it is dependent on how linear the

relationship is between a gas’s concentration and the signal it

produces in the cell (see

Fig.12.14

,

Fig.12.15a

and

Fig.12.15b

).

However, we can give you some guidelines on whether you have to

perform a single- or a two-point span-calibration.

·

Single-point calibration should be used when you wish to measure

in the linear range. We define the linear range as the dynamic

concentration range in which you can measure with an error of

5% of the real concentration (this is illustrated in

Fig.12.15b

). For

the most common gases, the linear range is from its detection

limit and up to 10 000 times its detection limit.

·

Two-point calibration is recommended to be used when you wish

to measure in the non-linear range and when you wish to

measure over a dynamic range of 100 000.

Gas Concentrations – Single-point Calibration

If a single-point span-calibration is performed then the span gas

should have a concentration which is at least 100 times its detection

limit and preferable not more than the highest concentration you

expect to measure.

For certain gases it can be inexpedient to use a concentration of at

least 100 times its detection limit, e.g. if you wish to measure a

poisonous gas with a high detection limit, it would be natural to use

a lower concentration. However, this means that the noise and

vibration will have a larger influence on the calibration. This must,

therefore, be taken into account when considering using a lower

concentration.

Gas Concentrations – Two-point Calibration

During a two-point span calibration, the total signal in the cell (V

low

)

is measured first with a low concentration (C

low

) of span gas, and

then the total signal in the cell (V

high

) is measured with a high

concentration (C

high

) of span gas. This enables the Monitor to

compensate for any un-linearity in the relationship between the

signal measured in the cell and the concentration of the gas in the

cell (this is illustrated, for a gas A1, in

Fig.12.15a

).

Here are some guidelines about the concentration of gas you should

use during a two-point span calibration:

·

The Low Concentration should be 10 to 30% of the highest

concentration you expect to measure.