4 setting up a monitoring task – LumaSense Technologies INNOVA 1412i User Manual
Page 42

Chapter 4
______________________________________________________________________
_____________________________________________________________________________
BE6025-15
1412i Photoacoustic Gas Monitor
LumaSense Technologies A/S
Page 42 of 202
f. Press
¿
.
g. Use the direction keys to define the tube flushing time.
Note: the range of acceptable values for tube flushing ranges
from 3s to 120s, increasing in 1s steps.
h. Press
¿
to select displayed flushing time.
The following text is displayed.
i. Press SET-UP
4.3.4
Setting up a Monitoring Task
The optical filters in the monitor have to be calibrated before they
can be used to measure gas concentrations. The user can only set-up
a monitoring task using those optical filters which have been
installed. In this example we have chosen to measure gas A (that is,
to measure gas A’s concentration using the optical filter installed in
position “A” of the filter carousel) and water vapour.
If your Monitor has only been calibrated to measure with another
filter (for example, the filter in position “B”) then choose gas B to
replace gas A in the following instructions. The parameters which
describe the monitoring task we shall set-up, are listed below:
TUBE FLUSHING TIME 3 s
PRESS ENTER TO CHANGE VALUE
NORMALIZATION TEMPERATURE 20.0°C
PRESS ENTER TO CHANGE VALUE