beautypg.com

1 producing a supply of clean, wet air – LumaSense Technologies INNOVA 1412i User Manual

Page 148

background image

Chapter 12

______________________________________________________________________

_____________________________________________________________________________

BE6025-15

1412i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 148 of 202

·

It allows you to have a visually check that there is a net flow of

gas out of the Teflon tube attached to it during the whole

calibration procedure. The “flow ball” can be seen to be bouncing

on the upward flow of air out of the meter.

If the pressure of the gas from the cylinder is too low at any time

then the “flow ball” will just remain seated at the bottom of its

tube. This condition must be avoided because atmospheric air will

be drawn into the Teflon tube system via the flow meter and

cause dilution of the calibration gas. This will affect the accuracy

of your calibration.

·

It functions as an escape valve. When the pressure of the gas in

the tubing becomes greater than atmospheric pressure, gas flows

out to the atmosphere via the flow meter. This ensures that the

gas entering the analysis cell is as close to atmospheric pressure

as possible, and this prevents damage to the very sensitive

microphones in the analysis cell.

WARNING! The analysis cell of the monitor is equipped with

sensitive microphones, and therefore NO direct connection between

the Monitor’s air-inlet and the pressure-valve on a gas cylinder is

allowed. Under NO circumstances should the pressure of the air in

the analysis cell be allowed to exceed 0.1 bar above the ambient

pressure.

1. Attach the Teflon tubing to the Air Inlet Filter at the back of the

monitor. This is described in detail in

section 4.2

.

2.

a. Connect the free-end of the length of tubing attached to the

air-inlet to one of the branches of the “Y”-piece (see

attachment point no. 1 in

Fig.12.5

).

b. Cut off another 1m length of Teflon tubing and connect one

of its ends to attachment point no. 2 of the “Y”-piece (see

Fig.12.5

) and its other end to a flow-meter.

3. Connect a 1m length of tubing to attachment point no. 3 of the

“Y”-piece (see

Fig.12.5

).

12.4.1 Producing a Supply of Clean, Wet Air

When humidity-interference calibrations are performed, clean, wet

air is required. This can be produced in the following manner.

We suggest that you bubble zero-gas (e.g. pure nitrogen) through a

thermostatically-controlled water-bath to produce a supply of clean,

wet air to the Monitor during humidity-interference calibration of the

filters (see

Fig. 12.7

). Notice that the thermostatically-controlled