Chapter 2
______________________________________________________________________
_____________________________________________________________________________
BE6025-15
1412i Photoacoustic Gas Monitor
LumaSense Technologies A/S
Page 18 of 202
Baud rate
9600
Stop bits
1
Data bits
7
Parity
Even
Hardwire mode
Leased line
Handshake type
Hardwire
Text Line
Terminator
CR-LF
Performing a full reset will set all parameters to the default settings.
2.4.2
Checking/Changing the RS-232 Communication
Parameters
The communication parameters for the serial interface must be set
using the push-buttons on the front of the monitor. An example is
given below
1. Press SET-UP S3 S1 S3 S1 .The screen display now shows the
following text.
If the baud rate displayed is incorrect press
¿
and use ▲ and ▼ to
display the correct value. Press
¿
again to store the selection.
If the baud rate displayed is correct, then press ▼ to continue to the
next parameter.
Press S1 to select 1 STOP BIT.
Press S1 to select 7 DATA BITS
Press S2 to select EVEN PARITY
Press S3 to select LEASED-LINE
Press S3 to select HARD-WIRED HANDSHAKE.
Press SET-UP to exit the set-up mode
Press RESET and S1 in order that the new settings are enabled.
2.4.3
Checking/Changing the Text line Terminator.
The Text line Terminator must be set using the push-buttons on the
front of the monitor.
Press SET-UP S2 ▼ ▼ ▼ S3
Press SET-UP to exit the set-up mode