81

HARDWARE ERR

10

An output error has occurred in an unknown output. Service is required.

HDW ERR CH 1

11

Errors 11 through 14 refer to a specific output where there is an output error.
Service is required.

HDW ERR CH 2

12

Same as in Error #11

HDW ERR CH 3

13

Same as in Error #11

HDW ERR CH 4

14

Same as in Error #11

NO MODEL NUM

15

The interface cannot find its model number. There may be a hardware failure
or the instrument may require reprogramming. Service is required.

CAL ERROR

16

You tried to use either a calibration command with CMODE off or the
calibration failed while in CMODE. Enable CMODE and check numbers sent
during calibration. Also, there could be a hardware error.

UNCALIBRATED

17

There is an incorrect checksum in the EEPROM possibly as a result of
incorrect calibration procedure. Recalibrate, and if the problem persists, your
supply has a hardware failure.

Table 5-8. Error Messages (continued)

Front Panel

Response

HP-IB

Code

Explanation

CAL LOCKED

18

Calibration was attempted with the Calibration Jumper on the HP-IB board in
the lockout position. Reposition jumper if desired. See Service Manual.

SKIP SLF TST

22

The self test jumper on the HP-IB board is in the Skip Self Test position. No
self-test was done. This is for diagnostics only. See Service Manual.

Table 5-9. TEST? Responses

Code

Explanation

0

This is the response to the TEST? query when there are no errors.

20

The timer has failed self-test. Refer to the troubleshooting section in the Service Manual.

21

The RAM has failed self-test. Refer to the troubleshooting section in the Service Manual.

27

The ROM has failed the checksum test. Refer to the troubleshooting section in the Service Manual.