// turn off autorange, set a 30V DCV range

errStatus = hpe1410_voltDcRang(vi, 0, 30);

check(vi, errStatus);

// set a 10 us aperture time

errStatus = hpe1410_voltDcAper(vi, 10.0e-6);
check(vi, errStatus);

// set 8000 readings
errStatus = hpe1410_sampCoun(vi, 8000);
check(vi, errStatus);

// store the readings in HP E8491A shared memory

// specify the E8491A shared memory base address
errStatus = hpe1410_memVmeAddr(vi, start_addr);
check(vi, errStatus);

// specify the amount of memory required

// (8000 readings * 8 bytes/reading)

errStatus = hpe1410_memVmeSize(vi, 64000);
check(vi, errStatus);

// enable the readings to be stored
errStatus = hpe1410_memVmeStat(vi, 1);

// disable multimeter error detection
hpe1410_errorQueryDetect(vi, 0);

// initiate the multimeter to take the readings
errStatus = hpe1410_initImm(vi);

// pause 30s to allow readings to complete
Sleep (30000);

// transfer the 8,000 readings (64,000 bytes) from the multimeter using
// the VISA function viMoveIn8
err = viMoveIn8(fw, VI_A24_SPACE, 0, 64000, (ViPUInt8)rdgs);
if(err < VI_SUCCESS) err_handler(fw, err);

// swap the bytes once they are transferred from shared memory
for (i=0;i<7999;i++)
{

SWAP_FLOAT64(&rdgs[i]);

}

// print some readings to verify the transfer was successful
for (i=0; i<10; i++)
{

printf("%lf\n", rdgs[i]);

}

// close the device sessions
hpe1410_close(vi); // close E1410 session
viClose(fw); // close E8491 session