Figure 8-10. fault 8, low heater temp -12, Oxymitter 4000, Alarms o2 temp low – Emerson OXYMITTER 4000 User Manual
Page 116: Instruction manual
Instruction Manual
IB-106-340 Rev. 3.0
December 2003
8-12 Troubleshooting
Rosemount Analytical Inc. A Division of Emerson Process Management
Oxymitter 4000
LOI
KEYPAD
DIAGNOSTIC
ALARMS
TEST
POINTS
HEATER T/C
HEATER
O2 CELL
CALIBRATION
CALIBRATION RECOMMENDED
O2 CELL mV +
O2 CELL mv -
HEATER T/C +
HEATER T/C -
INC
INC
DEC
DEC
HIGH
GAS
LOW
GAS
CAL
TEST GAS +
PROCESS -
% O2
SW2
TP1
J1
TP2
TP3
RED
YEL
GRN
ORG
TP4
TP5
TP6
ON
37260026
Alarms
O2 Temp Low
Figure 8-10. Fault 8, Low Heater Temp
h. Fault 8, Low Heater Temp
Figure 8-10 shows the electronic assembly
for an Oxymitter 4000 with a membrane
keypad (upper view) and an Oxymitter 4000
with an LOI (lower view).
Membrane Keypad. When Fault 8 is de-
tected, the HEATER LED flashes four times,
pauses for three seconds, and repeats.
1. The low heater temperature alarm is
active when the thermocouple reading
has dropped below 28.6 mV.
2. If the thermocouple reading continues
to ramp downward for one minute and
does not return to the temperature set
point of approximately 29.3 mV, then
an Open Heater fault will be displayed.
3. Power down the electronics. Remove
the electronic assembly per paragraph
9-4d, Electronic Assembly Replace-
ment. Using a multimeter, measure
across the heater connector, J8.
4. If the heater is good, the reading will be
approximately 70 ohms. If the heater is
open, see paragraph 9-4h, Heater Strut
Replacement.
LOI. When Fault 8 is detected, the LOI dis-
plays the “O2 Temp Low” message. Refer
to the comments and procedures in para-
graphs 1 through 4 above.