Troubleshooting 8-1
Oxymitter 4000
SECTION 8
TROUBLESHOOTING
8-1 OVERVIEW
While the Oxymitter 4000 electronics provides a
significant number of diagnostic alarms to assist
in troubleshooting potential problems, it’s good
to place these alarms in perspective with re-
spect to the instrument’s operating principles:
a. When the Zirconium Oxide sensing cell is
heated to its setpoint [1357°F (736°C)], the
cell will generate a voltage that represents the
difference between the process O
2
% and the
reference O
2
% inside the probe (20.95% O
2
ambient air).
b. Test points are provided to read the raw milli-
volt value generated by the thermocouple that
controls both the cell temperature and the raw
cell signal.
c. The cell temperature at test points 3 and 4
should always be stable at approximately
29 to 30 millivolts, which represents the
736°C setpoint temperature.
d. When flowing calibration gasses, the raw cell
millivolt value at test points 1 and 2 should
represent the levels on the chart in Figure
8-1. Note that the raw cell millivolt value in-
creases logarithmically as the O
2
concentra-
tion decreases.
37260043
100
10
1
0.01
0.1
0
0
50
100
150
200
Concentration O (%)
2
EMF (mV)
O Sensor Performance at 736 C
2
O
O
2
%
100
20
15
10
9
8
7
6
5
4
EMF(mV)
-34
1.0
7.25
16.1
18.4
21.1
23.8
27.2
31.2
36.0
O
2
%
3
2
1
0.8
0.6
0.5
0.4
0.2
0.1
0.01
EMF(mV)
42.3
51.1
66.1
71.0
77.5
81.5
86.3
101.4
116.6
166.8
Figure 8-1. O
2
Sensor mV Reading vs. % O
2
at 736ºC (Reference Air, 20.9% O
2
)