Recording with the data logger, Delete measurement data within the data logger – Amprobe PQ55A Power-Analyzer User Manual
Page 22
Data Memory – Saving individual measurements
This functions allows saving maximum 99 individual measurements.
A) Saving measurements
1) Switch on the instrument
2) Perform the measurement
3)
Save the desired measurement pressing the “M/MEMORY” key (19).
☞
When saving a measurement, the symbol [M] is briefly displayed, the storage
place is indicated below.
B) Read out the measurement
1)
Press the “R/READ” key (20). The symbol [R] (5.43) is displayed.
2) The navigation keys “5“ und “6” (16) allow to select and display the
measurements, starting with the first storage place.
3)
Press the key
(16a) to leave the memory mode again.
C) Delete measurements
1) Switch off the instrument
2)
Press the “M/MEMORY” key (19) and keep it pressed.
3) Switch on the instrument.
4)
Keep the “M/MEMORY” key pressed until the symbol “CLr“ is briefly
indicated in the lower right corner of the display.
Recording with the Data Logger
This functions allows to record up to 21,000 measurement values by means of the
internal data logger. For evaluation purposes, the internally recorded data must
be transferred to a PC (please refer to page 32).
☞
To use the total data memory, delete the data memory for the data logger
before starting the recording.
Delete Measurement Data within the Data Logger
☞
The values within the data logger may only be deleted using the software
supplied (please refer to page 32).
22