Model 2700 transmitter terminals
Figure 4-10:
A
B
C
D
E
F
G
H
I
J
K
A.
Brown
B.
Violet
C.
Yellow
D.
Orange
E.
Gray
F.
Blue
G.
White
H.
Green
I.
Red
J.
Mounting screw
K.
Ground screw (black)
4.4
Wire the transmitter to the sensor using
shielded or armored cable
For ATEX installations, shielded or armored cable must be installed with cable glands, at
both the sensor and transmitter ends. Cable glands that meet ATEX requirements can be
purchased from Micro Motion. Cable glands from other vendors can be used.
CAUTION!
Keep cable away from devices such as transformers, motors, and power lines, which produce
large magnetic fields. Improper installation of cable, cable gland, or conduit could cause
inaccurate measurements or flow meter failure.
CAUTION!
Install cable glands in the 9-wire conduit opening in the transmitter housing and the sensor
junction box. Ensure that the cable drain wires and shields do not make contact with the
junction box or the transmitter housing. Improper installation of cable or cable glands could
cause inaccurate measurements or flow meter failure.
Mounting and sensor wiring for 9-wire remote installations
40
Micro Motion
®
Model 1700 and 2700