Sensor and transmitter terminals, 1 sensor and transmitter terminals – Emerson MICRO MOTION 1700 User Manual
Page 42

Sensor and transmitter terminal designations
Table 4-5:
Wire color
Sensor terminal
Transmitter terminal
Function
Black
No connection
0
Drain wires
Brown
1
1
Drive +
Red
2
2
Drive –
Orange
3
3
Temperature –
Yellow
4
4
Temperature return
Green
5
5
Left pickoff +
Blue
6
6
Right pickoff +
Violet
7
7
Temperature +
Gray
8
8
Right pickoff –
White
9
9
Left pickoff –
d. Tighten the screws to hold the wire in place.
e. Ensure integrity of gaskets, grease all O-rings, then replace the junction box and
transmitter housing covers and tighten all screws, as required.
4.3.1
Sensor and transmitter terminals
ELITE, H-Series, T-Series, and some F-Series sensor terminals
Figure 4-8:
D
I
H
F
E
A B C
G
A.
Violet
B.
Yellow
C.
Orange
D.
Brown
E.
White
F.
Green
G.
Red
H.
Gray
I.
Blue
Mounting and sensor wiring for 9-wire remote installations
38
Micro Motion
®
Model 1700 and 2700