3. The spectrum lines on the screen do not update for long periods:

(1) Check if the screen is locked, if so, press Esc to unlock.
(2) Check if the instrument is in Remote control.
(3) Verify if all the trigger conditions have been met and if there is a valid

trigger signal.

(4) Check if the analyzer is in Single sweep.
(5) Check if the current sweep time is too long.

4. Wrong measurement results or poor precisions:

To verify the system errors and view the measurement accuracies and
precision, refer to the tolerance specifications or characteristics from the

User’s Guide

. To get the characteristics listed in the

User’s Guide

, please:

(1) Check if all the peripheral equipment has been successfully connected

and is working normally.

(2) Review the signal under measure and set appropriate instrument

parameters.

(3) Make measurements under proper conditions, for example: warm-up the

instrument appropriately and operate within the specified environment
and temperature.

(4) Calibrate the instrument regularly to reduce or avoid errors that can

occur over time.

Press SystemInformationSystem Info to get the last

calibration time and decide if a new calibration is required. If you
need a specific calibration after the stated calibration period,
contact

RIGOL for service options.

The analyzer provides a self-calibration function. If required, press

SystemCalibrateSelf-Cal, and the instrument will do a

self-calibration once the operating temperature has changed.

Press SystemCalibrateCal Now to do a self-calibration

immediately.