beautypg.com

Specifications – RIGOL DS6000 Series User Manual

Page 62

background image

RIGOL

26

User’s Guide for RP7000 Series Active Probe

Specifications


Technical Specifications
Item

RP7150

RP7080

Bandwidth

>1.5GHz

>0.8GHz

Rise Time

<265ps

<465ps

System Bandwidth

1GHz (DS6104 or
DS6102)

0.8GHz (DS6104 or
DS6102)

Input Capacitance

<1pF

Input Resistance

50kΩ±2% Differential
25kΩ±4% Single-ended

Input Dynamic Range

±6.25V

Input Common mode
Range

±6.75V DC to 100Hz
±1.25V >100Hz

Common Mode
Rejection Ratio

>45dB@1MHz

DC Attenuation

10:1 ±2%

Zero Offset Error

[1]

<30mV before calibration
<5mV after calibration

Offset Voltage Range

±12V


Offset Accuracy

[1]

<3% of current range
before calibration
<1% of current range
after calibration

Input Noise

45mVpp

Propagation Delay

7ns

Max Input Voltage

30V Peak CAT I

[2]

Electrostatic Protection >8kV

This manual is related to the following products: