beautypg.com

RIGOL DS6000 Series User Manual

Page 59

background image

RIGOL

User’s Guide for RP7000 Series Active Probe

23

calibration will last for about 40 to 50 seconds. When probe calibration
finished, “Probe calibration finished!” or “Probe calibration failure!” is
displayed in the user interface of the oscilloscope. Note: to ensure the
calibration precision, the probe tips must be firmly connected to the
calibration board during the calibration.

Note: Relative specifications of the RP7000 series active probe depend on the
calibration operation. After the calibration is finished, the DC gain, offset
voltage zero and offset gain will be calibrated. User can query the information
about the manufacturer, model, serial number and the last calibration time of
the probe through CH1ProbeProbe Info.














This manual is related to the following products: