To replace probe accessories, To replace probe accessories ) – RIGOL DS6000 Series User Manual
Page 55

RIGOL
User’s Guide for RP7000 Series Active Probe
19
① Solder-in single-ended probe head (RP7-0203).
② 0.2mm Nickel Wire (RP7-0306).
When the points to be tested are widely spaced, the length of the nickel wire
will be increased. At this point, overshoot and ringing will occurs and the
high-frequency response will changes.
To Replace Probe Accessories
1. To replace the probe head
Take care not to damage the connecting part to avoid affecting the probe
performance when replacing the probe head.
Replacing Method:
① Disconnect the current probe head from the active probe amplifier.
② Push the new probe head into the active probe amplifier straightly.
When single-ended probe head is used, pay attention to their
polarities.
2. Replace the probe tip
The
probe tip and probe head are connected with screw thread, so please
note the screw rotation and strength when removing and installing the
probe tip.
3. Replace the single-ended ground collar
Make sure the single-ended ground collar is firmly connected to the
copper pipe to ensure the probe performance when replacing it.