beautypg.com

Quickguide blower door – Retrotec Blower Door with DM-2 User Manual

Page 2

background image

Turn [Auto Zero] “On” to

zero gauge (every 8 seconds).

This is normally left on

unless batteries are low.
Press [Device] until “Retrotec 1000” is

displayed for 1000 systems, “Retrotec

2000” for Q46 & Q56 systems, “Retrotec

3000SR” for QMG, Q4E & Q5E systems, or,

“Retrotec DU200” for DU200 Systems.

Other devices can be removed using the

Setup Menu.
Use Range Ring B for most houses, try “C8”

for tighter new houses, or “A” for looser

older houses, if required.

Press [Mode] to cycle through results.

Check the Door Fan Operation Manual to

see the results required for

your region.
Follow the QuickGuide DM-2 mark II to

eliminate unused Devices, Range

Configurations and Modes.

Press [@ Pressure] to display results

exactly at (@) a test pressure. Use

[Set Pressure], or the menu in [Setup], to

change the @ Pressure value displayed.

Set [Time Avg] to “10s”. Increase if the test

pressure fluctuates over 1 Pa.

Wait for twice the Time Average length

before taking a reading. e.g. Set

[Time Avg] to “10s”, then wait

for 20 seconds

before taking

a reading.

Tip: Use longer time averaging in

windy conditions.

620 CFM is the flow rate that would occur at

25 Pa, even though only 22 Pa was achieved.

Flow reads “TOO LOW” or “----” at test pressure?
If the test pressure has been reached, but

“TOO LOW” or “----” appears, the fan is running too

slowly to measure flow.

Add the next Low-Range Ring.

Change [Range Config] on the DM-2 to match.

Re-adjust speed.

Cannot achieve test pressure at full speed?
If fan reaches 100% speed before reaching the target

pressure:

Remove a Range Ring and try again.

Change [Range Config] on the DM-2 to match.

Check seals on all registers. Look for disconnected

ducts or ducts open to outdoors.

Press [@ Pressure] to get the gauge to calculate

what the flow would be at exactly 25 Pa.

Step 4:

Desired results not achieved?

QuickGuide Blower Door

See: QuickGuide DM-2 mark II Digital Gauge

Press [On] twice. Gauge will start at previous setup.

Press [Baseline]. Wait 20 seconds on a calm day and

60 seconds on a windy day, the press [Enter].

Uncover fan. Install Range Ring B.

Press [Range Config] to show “B” on display.

Gauge set up

Step 3:

Conduct depressurization

test, (CFM@50)

Disconnect Speed Control Cable, then adjust Speed

Control knob until pressure

is about 50 Pa.

Record results.

Page 2 of 4