beautypg.com

Retrotec DM32 User Manual

Page 11

background image

Page 11 of 16
©Retrotec Inc. 2013

6. Update parameters not available in gauge “Settings” menu

The “Settings” tab on the Configurator lets you update parameters not available to change in the
“Settings” menu on the gauge itself. To open the Configurator screen shown in Figure 14, click on the
[Settings] tab at the bottom of the Configurator window.

If your gauge loses the calibration of the touchscreen, and tapping on the screen does not seem to get
the gauge to respond, you can use the Configurator to run the “Re-calibrate touchscreen” routine to
bring the screen back. The same routine will be run if you can press the reset button (on the bottom of
the gauge) with a paper clip or similar small item.

The AutoZero interval can be disabled, or changed from the default 30 seconds if the user has specific
need for this. It is recommended to leave the AutoZero enabled and with its interval at the default
setting.

Figure 14: Set AutoZero parameters and Re-calibrate touchscreen from the Configurator