Release or

Press/Hold

Displays

variance

(0.6

o

)

*

Press/Hold

Press once

Displays

maximum

*

Press/Hold

Press once more

Displays

minimum

Thermocouple burnout

Check sensor/wiring

Non-volatile memory error

De-power briefly

Replace unit if it persists

(Press together to

clear alarm condition)

Alarm condition: Temp exceeded maximum

operating temp (62 C).

Heater shutdown

until manually reset. Typically indicates

relay or temperature control failure.

Replace relay or temp control if persists.

PRESS

To clean interior and exterior surfaces, use a damp cloth with or without an all-purpose cleaner. The acrylic door

should only be cleaned using a lint-free cloth, with or without water. Paper towels can mar the surface of the

acrylic door. Use of any commercial cleansers on the acrylic door will cause crazing and cracking of the surface of

the acrylic over time. Periodically, check the accuracy of the control's temperature display against a known

accurate or calibrated device. This should be done with an empty chamber after the set temperature becomes

steady (typically after 45 to 60 minutes). Calibrate the control in the control's functions menu, level 3 (see page 3).

This feature monitors the stability of the control during for any given length process. It will record and display: 1.

The total variation or spread between high and low, 2. The absolute or maximum high and 3. The absolute or

minimum low.

To start the tracking feature navigate to "CheK" prompt in menu level 3. Hold star key then up arrow key to select

ON. Return to temp display or control will auto-return in 60 seconds. The control will track the temperature

variation until "CheK" is turned off. Recorded readings are retained until next "CheK" ON.

You can view readings at any time during or after tracking feature has been turned off. But de-powering the unit

will reset "CheK" to OFF and "rEAd" to zero. To view readings navigate to "rEAd" prompt in menu level 3. Then:

TIP: To avoid erroneous tracking data from run-up temperatures or door openings, start the Tracking feature after

articles have been placed and temperature steadies at set point.

PAGE 4

Control prompts will only display when a fault or alarm condition exists.

Temperature Tracking Feature

Control Self Diagnostics

Maintenance / Control Calibration

Tech Support

Warranty