Specifications
Maintenance
2
3
4
5
6
1
Always cover your microscope with the dust cover when not in use.
When cleaning the lenses, use lens paper or a cotton swab dipped in lens cleaning
solution.
Excess oil should be cleaned off your 100x objective and stage at once. An alcohol pad is best for removing oil
from the stage and on the other metal parts, but is not recommended for use on the lenses. Use lens cleaning
solution and lens paper to clean off your objectives.
Dust in the nosepiece or ocular tubes should be blown out using filtered air. Canned air dusters work well for this job.
Whenever you remove an objective, we recommend that you place it back into the original plastic shipping vial until
ready to be placed back on the microscope.
To keep your microscope in top condition for years, LW Scientific recommends that you have the microscope
professionally serviced once a year.
Warning: The 40x objective is not sealed for oil immersion. Damage to the 40x objective due to oil
immersion is not covered under warranty.
Power
The circuit board, in the base of the unit, contains all the electrical functions. There are no user
repairable parts on the circuit board. If you suspect a faulty board, call LW Scientifics’ technical
service department.
INPUT: AC90-240/50-6- HZ
OUTPUT: 6V/30W to a quartz halogen lamp (LW Scientific replacement part number: R-6V/30W)
FUSE: One 2-amp fuse protects the circuit board from electrical overload. The fuse is located in the snap-out fuse panel
next to the on/off switch. There is also a spare fuse located inside the same panel. When replacing the fuse, always
install a new one of the same size and amperage.
Nosepiece
Quintuple nosepiece
Head
Illumination
6V/30W halogen bulb
Stage
Dimensions and Weight
Weight:24.0 lbs.
Trinocular head
Ideal for video/photography
Centering eyepiece for phase objective
EW 10X /22 extra wide field eyepieces
Stage, 160x250mm
Auxiliary stage, 70x180mm
Coaxial drive controls
Optional mechanical stage available:
X-Y Coaxial control
Range of traverse: 120x78mm
Objectives
Infinity Optical System
Objectives: (Plan achromatic) 4X and 40x,
(Plan phase contrast) 10x and 20x
Condenser
Abbe condenser with 20X phaser
P. 770.270.1394
F. 770.270.2389
865 Marathon Parkway Lawrenceville GA 30046
MKT-7.5.3-L-112 | Rev 0