beautypg.com

Specifications, Brush replacement – LW Scientific M24 User Manual

Page 4

background image

1

2

3

4

5

6
7

Specifications

P. 770.270.1394

F. 770.270.2389

865 Marathon Parkway Lawrenceville GA 30046

Brush Replacement

Unplug the centrifuge, and turn it on its side.

Remove the feet and all Phillips head screws holding

the metal base plate.

Using a flat head screwdriver, remove the two black

caps on the bottom part of the motor.

Using your fingers, pull the brush out of the recessed

hole that the black caps were removed from. The

brushes consist of springs with a graphite bar on them.

Place the new brushes in the recessed hole, and

replace the black caps (See image, right).

Replace the metal base plate and the feet.

Your LWS-M24 is now ready to use.

Nominal speed:

Max. Volume:

Input Voltage:

Height:

Width:

Depth:

Weight:

Boxed dimensions:

Boxed weight:

12,000 RPM (+/- 500)*

24 - 75mm capillary tubes

110V or 220V

8.7 inches

10.6 inches

12.5 inches

15.8 lbs

11.5 x 10 x 7 inches

23 lbs.

*These are average speeds, which may vary slightly between units and with load size. Please use a tachometer for actual

RPMs on your unit, and for calibration purposes.

MKT-7.5.3-L-032 | Rev 1