Specifications, Brush replacement – LW Scientific M24 User Manual
Page 4
1
2
3
4
5
6
7
Specifications
P. 770.270.1394
F. 770.270.2389
865 Marathon Parkway Lawrenceville GA 30046
Brush Replacement
Unplug the centrifuge, and turn it on its side.
Remove the feet and all Phillips head screws holding
the metal base plate.
Using a flat head screwdriver, remove the two black
caps on the bottom part of the motor.
Using your fingers, pull the brush out of the recessed
hole that the black caps were removed from. The
brushes consist of springs with a graphite bar on them.
Place the new brushes in the recessed hole, and
replace the black caps (See image, right).
Replace the metal base plate and the feet.
Your LWS-M24 is now ready to use.
Nominal speed:
Max. Volume:
Input Voltage:
Height:
Width:
Depth:
Weight:
Boxed dimensions:
Boxed weight:
12,000 RPM (+/- 500)*
24 - 75mm capillary tubes
110V or 220V
8.7 inches
10.6 inches
12.5 inches
15.8 lbs
11.5 x 10 x 7 inches
23 lbs.
*These are average speeds, which may vary slightly between units and with load size. Please use a tachometer for actual
RPMs on your unit, and for calibration purposes.
MKT-7.5.3-L-032 | Rev 1