Care and Maintenance

1

Use only high quality test tubes. Lower quality or inexpensive glass or plastic tubes may fracture and release

their contents into the tube chamber. Make sure you know the maximum force allowed for the tubes you are

spinning.

Never force a tube into the tube shield. The tube shields were design to accommodate most common sizes

of tubes.

Keep the tube shields clean. If a tube breaks inside a shield, clean all the debris from the shield and bowl and

disinfect.

If a large amount of fluid has spilled inside the unit, carefully remove the tube shields and rotor and clean the

inside of the rotor chamber and the top of the motor.

*The horizontal metal rotor and tube shields can be autoclaved.

3

4

2

Specifications

MKT-7.5.3.-L-110 | Rev 6

Nominal Speed:

500-3400 rpm

Fuse: 10A

Height: 9.9in

Width:

14.9in

Depth: 14.5in

Weight: 12lbs

Boxed Dims:

16x16x12in

Boxed Weight: 17lbs

Fluid

Recommendations: Speed

Time

Whole Blood

3400

10 min.

Urine

1800

5-10 min.

The Universal Centrifuge is intended for use as a general laboratory centrifuge.

LED Display Definitions:

Green LED ON

Lid is locked / Ready for cycle

Green LED blinking quickly

RED LED ON
Green LED blinking slowly

RED LED ON
Red LED blinking/display is “00” Lid locked / Cycle complete

Unit accelerating

Unit at operating speed

RCF

(g’s)

31
80

125
180
245
282
321
406
501
606
721
846
982

1127
1282
1364

Radius

(mm)

112
112
112
112
112
112
112
112
112
112
112
112
112
112
112
112

RPM

500
800

1000
1200
1400
1500
1600
1800
2000
2200
2400
2600
2800
3000
3200
3300

G-Force with

Fixed Rotor:

RCF

(g’s)

39

100
156
225
306
352
400
507
626
757
901

1058
1227
1408
1602
1704

Radius

(mm)

140
140
140
140
140
140
140
140
140
140
140
140
140
140
140
140

RPM

500
800

1000
1200
1400
1500
1600
1800
2000
2200
2400
2600
2800
3000
3200
3300

G-Force with

Swing-Out Rotor:

1450 112

3400

1809 140

3400