beautypg.com

Model nga 2000 hfid – Emerson Process Management 760004-A User Manual

Page 9

background image

Instruction Manual

760004-A

February 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3

Model NGA 2000 HFID

WARNING

.

HIGH TEMPERATURE

This equipment is used in the analysis of sample gases at temperatures of up to 250

C. All compo-

nents and material in contact with the sample, the oven and the burner can reach this temperature
level.

Operate this equipment only when covers are secured. Servicing requires access to "hot" parts
which can cause serious injury. Refer servicing to qualified personnel.

NOTE

This Analyzer Module is completely leak-tested at the factory for gas leakage. The user is responsi-
ble for testing for leakage at the inlet and outlet fittings on the rear panel (with a test procedure cho-
sen by the user). The user is also responsible for leak-testing periodically and if any internal
pneumatic components are adjusted or replaced. See leak test instructions in Section 2-5.

WARNING

.

PARTS INTEGRITY

Tampering with or unauthorized substitution of components may adversely affect safety of this
product. Use only factory-approved components for repair.

CAUTION

.

PURGE AIR REQUIREMENT

This Analyzer Module must be used in conjunction with a device (Control Module or PC Interface)
that can actively monitor network variables related to pressure or flow of the continuous dilution
purge, or the front panel LEDs of the Analyzer Module, as installed, must be visible. The purpose of
this requirement is to maintain adherence to ANSI/NFPA 496 standard which assures the continued
viability of the purge system. Under no circumstances should any pressure or flow indicator be
connected to the PURGE AIR OUT outlet of the Analyzer Module because this may affect the sealing
performance of the module.

CAUTION

.

PRESSURIZED GAS

This module requires calibration with a known standard gas. See General Precautions for Handling
and Storing High Pressure Gas Cylinders, page P-5.