Instruction Manual

760004-A

February 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Maintenance and Service 4-1

Model NGA 2000 HFID

Cover

Insulation Shield

Securing
Screw (4)

Cover/Rear Panel Secur-
ing Screws (6)

ATTE

NTION

PURGE

AIR OUT

WARNING

WARNING

ATTENTION

BURNER

AIR IN

REGULATED

AIR IN

PURGE

AIR IN

HEATED

SAMPLE
BYPASS

HEATED

SAMPLE IN

BURNER
EXHAUST
OUT

FUEL
IN

PRESSURE
RELIEF
VALVE

Front Panel Securing
Screws (6)

SECTION 4

MAINTENANCE AND SERVICE

WARNING

QUALIFIED PERSONNEL

This equipment should not be adjusted or
repaired by anyone except properly quali-
fied service personnel.

4-1 OVERVIEW

This section contains instructions and proce-
dures for troubleshooting and maintaining the
HFID analyzer module. To access the inter-
nal components of the analyzer module, per-
form the following:

1. Remove power to the unit; shut off gases

and disconnect lines. Allow module to
cool.

2. Refer to Figure 4-1. Remove the six

screws securing the front panel, then the
six screws securing the cover to the rear
panel. Slide cover towards rear panel to
remove. Loosen four screws securing in-
ner insulation shield to base, lift up to re-
move.

Figure 4-2 illustrates the locations of major
components of the HFID.

Figure 4-1. Removal of Cover and Insulation Shield