beautypg.com

5 specifications, A. general, B. physical – Emerson Process Management 760004-A User Manual

Page 18: Model nga 2000 hfid, Instruction manual

background image

Instruction Manual

760004-A
February 2002

1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA 2000 HFID

1-5 SPECIFICATIONS

a. General

Measurement Species ....................Total hydrocarbons
Ranges (H2/He Fuel)

Low range .............................. 0 to 10 ppm, CH4, through 0 to 1%, CH4 at an oven setpoint be-

tween 113

°C and 191°C

High range ............................. 0 to 50 ppm, CH

4

, through 0 to <5%, CH

4

at an oven setpoint be-

tween 113

°C and 191°C

Analysis Temperature .....................Adjustable from 200

o

F to 400

o

F (93

o

C to 204

o

C), maintained within

±11

o

F (±6

o

C) from the setpoint.

Repeatability ....................................

≤1% of fullscale for successive identical samples, at a constant
temperature, sample flow and fuel, burner air, regulated air and
sample pressures

Min. Detectable Level......................0.10 ppm, CH

4

Noise ................................................

≤1% of fullscale, peak to peak

Linearity............................................

≤±1% of fullscale, ≤±2% of data point (must be above the minimum
detectable level)

Response Time ...............................

≤1.5 sec., 0% to 90% of fullscale

Drift

Zero .......................................

≤ ±1% of fullscale/24 hours at constant temperature, sample flow,

hydrocarbon concentration of supply gases, and
fuel, burner air, regulated air and sample pres-
sures.

Span ......................................

≤ ±1% of fullscale/24 hours at constant temperature, sample flow,

hydrocarbon concentration of supply gases, and
fuel, burner air, regulated air and sample pres-
sures.

Effect of Temperature....................

≤±2% of fullscale for any ambient temperature change of 10

o

C and

rate of change less than 10

o

C/hr.

Operating Temperature ................. 59

°F to 95°F (15°C to 35°C)

Power Requirements.......................+24 VDC ±5%, 120 W max. direct to Analyzer Module

Ripple and Noise: <100 mV pp
Line and Load Regulations: <1%

b. Physical

Case Classification........................ General purpose for installation in weather-protected area
Maximum Separation .................... 1600m (1 mile) from Analyzer Module to Platform
Materials in Contact With Sample . Stainless steel and glass-filled Teflon
Dimensions.................................... See Outline and Mounting Dimensions, Figure 2-5
Weight ........................................... 15.9 kg (35 lbs.)
Mounting........................................ Horizontally, custom-installed in a panel