Custom curves, Calibration, Custom curves calibration – Super Systems DAQ User Manual
Page 23

Super Systems Inc. Page 23 of 34
DAQ Operations Manual
Custom Curves
The
Custom Curves
menu option will allow the user to set up the variables for the curves.
Curves are applied when the user wants to use a non-linear compensation to the input, such as
in vacuum areas. There are a maximum of seven curves.
The Interpolation Type can be one of the following:
None
Linear
The mV 1 through mV 32 can be within the range of –32768 to 32767.
The Offset 1 through Offset 32 can be within the range of –32768 to 32767.
Note: This is where the user will create the custom curve that can be applied to the inputs
through the
Range Setups
menu option
. See the
section for information on how
to apply a curve to an input.
An example of a non-linear curve is for a 0 to 1V span (0 to 1000 mV). If the mV is 0, then the
offset is 0. If the mV is 200, then the offset if 400, etc.
Calibration
The
Calibration
menu option will allow the user to calibrate the inputs on a board. Click on the
click value to begin the calibration process.