Seaward DP6 User Manual
Page 11
Operating Instructions
Thermocouple Calibrator Type DP6
10
The following may be changed :
1. The units °F °C °K mV V
2. Source or Sense
3. Thermocouple Type or Voltage
4. Cold Junction - automatic/manual/off
RIGHT HAND MENU DISPLAY
To change the current configuration, press the SEL (Select) Key. The first
changeable parameter will flash. To change this, press the CHG (Change) key and
another menu will appear.
If you do not wish to change the flashing parameter, press SEL again and the next
variable will flash. Repeated pressing of the SEL key will scroll through the possible
changeable characters.
Once the CHG key is pressed and the second menu appears, a number of choices
will be possible. The options will be directly above the white arrows. Choose the
option you require and follow the arrow down, pressing the key directly below it. In
some cases, a third option menu will appear giving more choice of configuration.
EXAMPLE:
To configure the DP6 in SOURCE mode for a type K thermocouple with cold junction
OFF, proceed as follows:
Press SEL - measure/source flashes
Press CHG - second menu appears
Press the key below SOURCE (KYB) DP6
switches to source mode and next changeable
parameter (thermocouple type/DCV) flashes
Press CHG - next menu appears
Press key below T/C (SEL) thermocouple choice
menu appears
Select thermocouple type with flashing cursor
using the keys below « » symbols. Once the
cursor is placed on the required thermocouple
type, press OK key.