beautypg.com

Caution – Measurement Computing WBK40 User Manual

Page 9

background image

DBK84

DBK84

DBK84

WBK40


WaveBook/516E

WBK17

WBK10A

Note 1

Channel Input Side Device Interface Side

A WaveBook/516E with a WBK17, WBK10A, WBK40, and three DBK84 Modules

Device Interface Connections (seen in the above example):

1. From the WBK40 P1 connector to three DBK84 Modules via three CA-37-1T cables.*
2. From the WBK40 SYNC port to WaveBook/516E SYNC Port via a CA-74-1 cable.
3. From the WBK40 PARALLEL PORT to the WaveBook/516E Expansion Port via a CA-35-12 cable.
4. From the WaveBook/516E’s ETHERNET port to the Ethernet via a CA-242 cable.
5. From the WaveBook/516E EXTERNAL CONTROL connector to the WBK17’s EXPANSION

CONTROL IN connector, via a CA-129 cable.

6. From the WaveBook/516E EXTERNAL SIGNAL IN connector to the WBK17 EXTERNAL

SIGNAL OUT connector via a CA-150 cable.

7. From the WaveBook/516E Channel Input Side’s DIGITAL I/O AND TRIGGER connector to the

WBK17 DIGITAL EXPANSION IN connector via a CA-217 cable.

8. From the WBK17 EXPANSION CONTROL OUT connector to the WBK10A EXPANSION

CONTROL IN connector via a CA-129 cable.

9. From the WBK17’s EXPANSION SIGNAL IN connector to the WBK10A’s EXPANSION SIGNAL

OUT connector via a CA-150 cable.

10. Power Supply – a TR-40U was connected to the POWER IN [DIN 5] connector to supply power at

+10 VDC to +30 VDC for the WBK40, WaveBook/516E, and WBK17.

11. Power Supply – From the WBK17 POWER OUT [DIN 5] +10 VDC to +30 VDC connector to the

WBK10A POWER IN [DIN 5] connector via a CA-115 power cable. The CA-115 is a 6-inch long
cable which has a 5-pin DIN male connector at each end.

CAUTION

An incorrect use of power can damage equipment or degrade performance. Prior to
connecting your devices to power, calculate your system’s power requirements.

Reference Note
Information pertaining to connecting the system to power and calculating the amount of
power needed can be found in the WaveBook User’s Manual. A PDF version is included on
the data acquisition CD and can be accessed from the CD’s intro-screen by using the PDFs> button.

* Information regarding the use of CA-37-T cables is provided on page 10.

Using Fastener Panels to Stack Modules:

Fastener panels are illustrated and discussed on page 8.

Note 1: Each DBK84 module requires a unique address setting as explained on page 13.

Thermocouple & Multifunctional Modules

926896

WBK40 & WBK41, pg. 9

This manual is related to the following products: