Specifications – WBK40 and WBK41

Power

Supply Voltage Range: 10 to 30 VDC
Power Out, from P1:

+15 VDC @ 500 mA (via pin 21)
–15 VDC @ 500 mA (via pin 2)
+5 VDC @ 1000 mA (via pin 1)

Up to eight DBK84 options can be powered from the DB37 P1 connector.
For applications with 9 to 15 DBK84 options, a DBK32A must be added to the system.
The WBK41 includes a Power Terminal Block on the front panel. When used, it draws
from the total power available to P1.

Fuses:

Input Power Fuse (F201): 4A MINI ATO, user replaceable; refer to page 23 for details.
5V Power Output Fuse (F1): 2.5A resettable; not user replaceable.

General

Operating Temperature: 0° to 50°C (32° to 122°F)
Storage Temperature: -40° to 70°C (-40° to 158°F)
Relative Humidity: 0 to 95%, non-condensing
Vibration: MIL STD 810E
System Connectors:

Analog Expansion: P1, DB37 male connector on rear panel.
Digital Expansion: P2, DB37 male connector on WBK41 rear panel. P2 does not apply to WBK40.

TC/mV Connector: Mini-TC connectors (WBK40 & WBK41)
Digital I/O, Power [out], Counter/Timer, DAC: Removable Screw-Terminal Blocks on front panel, WBK41 only.

A factory installed DBK46 option is required for DAC outputs (see page 17).

Dimensions: 285 mm W x 220 mm D x 70 mm H (11” x 8.5” x 2.70”)
Weight: 1.6 kg (3.5 lbs)

A/D Specifications

Type: Successive approximation
Resolution: 16 bit
Conversion Time: 5 µs
Maximum Sample Rate: 200 kHz
Nonlinearity (Integral): ±1 LSB
Nonlinearity (Differential): No missing codes

Thermocouple Inputs

Functions: TC types J, K, S, T, E, B, R, N; x100 (voltage)
Inputs: 14 differential TC/mV inputs
Input Voltage Range: ±100 mV
Input Impedance: 40M Ohm (differential); 20M Ohm (single-ended)
Input Bandwidth: 4 Hz
Input Bias Current: 10 nA typical
CMRR: 100dB typical
Maximum Working Voltage (signal + common mode): ±10V
Over-Voltage Protection: ±40V
Voltage Accuracy: ±(0.2% of reading + 50 µV)
TC Accuracy: See the following table. Valid for one year, 18 to 28°C
Minimum Resolution: 0.1°C for all TC types

Note: Specifications are subject to change without notice.