beautypg.com

Measurement Computing WBK40 User Manual

Page 29

background image

TC Accuracy at Measurement Temperature in °C (±°C)

Type

Min

Max

-100

0

100

300

500

700

900

1100

1400

J

-200

760 0.6 0.5 0.5 0.6 0.7 0.7 —

— —

K

-200

1200

0.7 0.6 0.6 0.7 0.8 0.8 0.9 1.0 —

T

-200

400 0.7 0.6 0.6 0.6 — — —

— —

E

-270

650 0.6 0.5 0.5 0.5 0.6 — —

— —

S

-50

1768

— 2.3 1.8 1.5 1.5 1.4 1.5 1.6 1.6

R

-50

1768

— 2.3 1.6 1.5 1.4 1.4 1.3 1.4 1.5

B

50

1780

— — — 3.7 2.4 2.1 1.8 1.7 1.5

N28

-270

400 0.9 0.7 0.7 0.7 — — —

— —

N14

0 1300

— 0.7 0.7 0.7 0.8 0.8 0.9 1.0 —


Input Sequencer

Scan Clock Sources: 2
1. Internal, programmable from 5 µs to 5.96 hours in 5 µs steps
2. External, TTL level input up to 200 kHz max; pin 20 on P1.
Depth: 512 locations
Channel-to-Channel Scan Rate: 5 µs per channel, programmable
Data Buffer: 64 Ksample FIFO
Note for WBK41 Only: Analog, digital and frequency inputs can be scanned synchronously,

based on either an internal programmable timer, or an external clock source.

External Acquisition Scan Clock Input

Maximum Rate: 200 kHz
Clock Signal Range: 0V to +5V
Minimum Pulse Width: 50 ns high, 50 ns low
External SYNC Port: Available on rear panel, allows the modules to be synchronized with a WaveBook/516E

Note: Specifications are subject to change without notice.

Thermocouple & Multifunctional Modules

926896

WBK40 & WBK41, pg. 29

This manual is related to the following products: