beautypg.com

Tb - ieee bus terminator – Measurement Computing Micro488/p User Manual

Page 31

background image

Command Descriptions

09-27-02

3-13

TB

IEEE BUS TERMINATOR

The IEEE BUS TERMINATOR command is used to select the IEEE bus terminator for
reads from the IEEE bus. The selected terminator signifies the end of an ENTER sequence.
The available terminators are:

1 - LF
2 - CR
3 - LF-CR
4 - CR-LF

SYNTAX

TB;n

n is one of the available terminator types.

RESPONSE

None

BUS STATES

None

EXAMPLE

PRINT #1, "TB;1"

Select LF as an IEEE bus
terminator.