Analog input/output calibration – Measurement Computing USB-2408 Series User Manual
Page 27

USB-2408 Series User's Guide
Specifications
27
Parameter
Conditions
Specifications
Power on and reset state
DACs cleared to zero-scale: 0 V,
±50 mV
Output noise
60 µV
rms
(BW=1.5 KHz)
Settling time
To rated accuracy, 10 V step
75 µS
Slew rate
1.0 V/µs
Throughput
Single-channel
1000 S/s max, system-dependent,
0.01 S/s min
Multi-channel
1000 S/s / #ch max, system-
dependent, 0.01 S/s min
Table 14. Calibrated absolute accuracy specifications
Range
Accuracy (±LSB)
±10 V
16.0
Table 15. Calibrated absolute accuracy components specifications
Range
% of reading
Offset (±mV)
Temp drift (%/°C)
Absolute accuracy at FS (±mV)
±10 V
±0.0183
1.831
0.00055
3.661
Table 16. Relative accuracy specifications
Range
Relative accuracy (±LSB)
±10 V
4.0 typ
Analog input/output calibration
Table 17. Analog input/output calibration specifications
Parameter
Specifications
Recommended warm-up time
45 minutes min
Calibration
Firmware calibration
Calibration interval
1 year
AI calibration reference
+10.000 V, ±5 mV max. Actual measured values stored in EEPROM
Tempco: 5 ppm/°C max
Long term stability: 30 ppm/1000 hours
AO calibration procedure
(USB-2408-2AO only)
The analog output pin is internally routed to the analog input pin.
AOUTx readback
(USB-2408-2AO only)
Each AOUTx output can be independently measured by the onboard A/D converter.
Software-selectable