System throughput, Settling time – Measurement Computing PCI-DAS6013 User Manual
Page 32

PCI-DAS6013 and PCI-DAS6014 User's Guide
Specifications
32
Table 2. Absolute accuracy components specifications - All values are (±)
Range
% of
Reading
Offset
(µV)
Averaged Noise +
Quantization (µV)
1
Temp Drift
(%/DegC)
Absolute
Accuracy at FS
(mV)
±10 V
0.070
1800
180
0.001
8.984
±5 V
0.020
918
85
0.001
2.003
±500 mV
0.070
109
12
0.001
0.471
±50 mV
0.070
27
7
0.001
0.069
1
Averaged measurements assume averaging of 100 single-channel readings.
Each PCI-DAS6013 and PCI-DAS6014 is tested at the factory to assure the board’s overall error does not
exceed accuracy limits described in Table 1.
Table 3. Differential non-linearity specifications
All ranges
±0.5 LSB ty,p ±1.0 LSB max
No missing codes
16 bits, guaranteed
System throughput
Condition
Calibration Coefficients
ADC Rate (max)
1. Single channel, single input range
Per specified range
200 kS/s
2. Multiple channel, single input range
Per specified range
200 kS/s
3. Single channel, multiple input ranges
Default to value for cbAInScan() range parameter
200 kS/s
Note 2:
For conditions 1-2 above, specified accuracy is maintained at rated throughput. Condition 3
applies a calibration coefficient which corresponds to the range value selected in
cbAInScan()
.
This coefficient remains unchanged throughout the scan. Increased settling times may occur
during gain-switching operations.
Settling time
Settling time is defined as the time required for a channel to settle to within a specified accuracy in response to
a full-scale (FS) step. Two channels are scanned at the specified rate. A –FS DC signal is presented to
channel 1; a +FS DC signal is presented to channel 0.
Condition
Range
Accuracy
±0.0031%
(±2.0 LSB)
±0.0062%
(±4.0 LSB)
Same range to same range
±10 V
5 µs typ
*
±5 V
5 µs max
*
±500 mV
5 µs typ
*
±50 mV
*
5 µs typ