The overall accuracy of any instrument is limited by the error components within the system. Quite often,
resolution is incorrectly used to quantify the performance of a measurement product. While "12-bits" or "1 part
in 4096" does indicate what can be resolved, it provides little insight into the quality of an absolute
measurement. Accuracy specifications describe the actual results that can be realized with a measurement
device.
There are three types of errors which affect the accuracy of a measurement system:
!
offset
!
gain
!
nonlinearity
The primary error sources in the miniLAB 1008 are offset and gain. Nonlinearity is small in the miniLAB 1008,
and is not significant as an error source with respect to offset and gain.
Figure 3-9. Ideal ADC transfer function
shows an ideal, error-free, miniLAB 1008 transfer function. The typical calibrated accuracy of the
miniLAB 1008 is range-dependent, as explained in the "
" chapter of this document. We use a ±10
V range here as an example of what you can expect when performing a measurement in this range.
-FS
+FS
Input Voltage
0
4095
Output Code
2048
The miniLAB 1008's offset error is measured at mid-scale. Ideally, a zero-volt input should produce an output
code of 2048. Any deviation from this is an offset error.
shows the miniLAB 1008 transfer function
with an offset error. The typical offset error specification on the ±10 V range is ±9.77 millivolts (mV). Offset
error affects all codes equally by shifting the entire transfer function up or down along the input voltage axis.
3-10