beautypg.com

6 refractometer difference logic – K-Patents DD-23 User Manual

Page 34

background image

30

7.6 Refractometer difference logic

If there are two refractometers in operation (taken into operation, not washing), the
Control unit monitors the difference in reading between the two instruments (󰅮ig-
ure 7.6). If the difference is more than 2%, a refractometer difference warning is given
with the corresponding front panel indicator and relay.

The divert decision logic (section 7.2) is not affected by the refractometer difference,
but a signi󰅮icant difference between the two instruments indicates one of the instru-
ments is not measuring reliably. In this case the instruments and wash systems have
to be checked immediately. Ignoring to do this may result in an erroneous divert de-
cision.

no REFRACTO-

METER

DIFFERENCE

|A-B| > 2%?

2 instruments

available?

yes

no

no

yes

REFRACTO-

METER

DIFFERENCE

Figure 7.6

Refractometer difference logic