6 refractometer difference logic – K-Patents DD-23 User Manual
Page 34

30
7.6 Refractometer difference logic
If there are two refractometers in operation (taken into operation, not washing), the
Control unit monitors the difference in reading between the two instruments (ig-
ure 7.6). If the difference is more than 2%, a refractometer difference warning is given
with the corresponding front panel indicator and relay.
The divert decision logic (section 7.2) is not affected by the refractometer difference,
but a signiicant difference between the two instruments indicates one of the instru-
ments is not measuring reliably. In this case the instruments and wash systems have
to be checked immediately. Ignoring to do this may result in an erroneous divert de-
cision.
no REFRACTO-
METER
DIFFERENCE
|A-B| > 2%?
2 instruments
available?
yes
no
no
yes
REFRACTO-
METER
DIFFERENCE
Figure 7.6