Ivoclar Vivadent IPS e.max Abutment Solutions Cem Kit v.3 User Manual
Page 12
12
Apply Multilink Implant directly from the mixing tip in a thin layer on the
bonding surface of the Ti base.
Position the meso-structure above the Ti base in such a way that the position
markings are aligned.
Carefully remove excess in the screw cavity, e.g. with a Microbrush or a
brush, using a rotary movement.
Apply a glycerine gel (e.g. Liquid Strip) to the cementation joint to prevent
the formation of a inhibition layer.
Polymerize the luting composite for 7 minutes (optionally in a light-curing
unit). Important: Do not move the objects until the material has
completely cured and hold them in place without allowing
for any motion.
Apply Multilink Implant directly from the mixing tip in a thin layer on the
bonding surface of the meso-structure.
Join the components using even and light pressure. Subsequently, tightly
press the components together for 5 seconds.
Remove excess carefully in its ductile state, e.g. with a foam pellet, while
applying slight pressure to hold the components in place.