5. Operation and Configuration

3. Load the furnace.
You can now load the firing chamber with the hot investment ring.
The IRT investment ring analysis starts immediately after loading of
the furnace:

– automatic identification of the investment ring size
– check of the investment ring preheating temperature.

For details regarding the IRT investment ring analysis see
Chapter 5.4.

2. Select the program.

4. Start or edit the press program.
Now, the press program can be started. Individual press programs
can be edited.

5.3 Press programs and programming options

The furnace offers different press program ranges:

a. press programs for Ivoclar Vivadent materials
b. 20 free, individually adjustable press programs.

a) Press programs for Ivoclar Vivadent materials
When the furnace is delivered ex factory, the Ivoclar Vivadent programs already contain the recommended material
para meters. They cannot be deleted or overwritten. The parameters are not shown in the software.

b) Free, individually adjustable press programs
Individual press programs can be freely programmed and renamed.

In case of software updates, individually changed parameters in free programs are NOT reset to the factory

settings or changed!

5.3.1 Program selection in the "Pressing" mode and with activated IRT investment ring analysis
The program selection requires only a few steps:

1. Select the program type.