Maintenance, cleaning, diagnosis – Ivoclar Vivadent EP 600 Combi User Manual

Page 27

background image

27

What

Part

When

Check all plug-in connections for correct fit

Various external connections

weekly

Check if the furnace head opens smoothly and without

Furnace head opening mechanism

monthly

excessive noise

Check if the thermocouple is straight and in the right place

Thermocouple

weekly

Check the stone lining inserts for large cracks or damage.

Stone lining inserts

weekly

If the stone linings are worn down, they must be replaced
by a certified Ivoclar Vivadent Service Center. Slight
hairline cracks on the stone lining surface do not have a
negative influence.

Check if the sealing rims of the furnace head and furnace

Sealing rims of the furnace head

weekly

base are clean and undamaged

and furnace base

Check the keypad for visible damage. If the keypad is

Keypad

weekly

damaged, it must be replaced by a certified
Ivoclar Vivadent Service Center.

Check furnace temperature (calibration).

Press chamber

Every 3–4 months, or after
approx. 50 press cycles.

7.2 Cleaning

The furnace may only be cleaned when
it is cool, since there is a burn hazard.
Do not use any cleaning solutions.

Parts

Frequency

Cleaning material

Housing

if required

soft, dry cloth

Keypad

weekly, or if required

soft, dry cloth

Tray

if required

cleaning brush or vacuum cleaner

Stone lining inserts

daily

cleaning brush

Sealing rims of the furnace head and

daily

cleaning brush and a soft cloth

furnace base

7. Maintenance, Cleaning, Diagnosis

This chapter describes the
maintenance and cleaning
procedures that may be
performed by the users. All
other tasks must be performed
by qualified service personnel at
a certified Ivoclar Vivadent
Service Center.

7.1 Monitoring and
maintenance

The time for these maintenance
procedures depends on the
frequency of use and the
working habits of the users. For
that reason, the recommended
times are only approximates.

Turn off the furnace
and disconnect power
before maintenance
and cleaning, since
there is a risk of
electrical shock.

This apparatus has

been developed for
typical use in the
dental laboratory.

If the product is used in a pro-
duction facility, for industrial
applications, or in continuous
firing operation, premature
ageing of certain spare parts
have to be expected.

These spare parts are e.g.:
– Heating muffle
– Insulation material
– Lamps

These spare parts are not
covered by the warranty.

Please also observe the shorter
service and maintenance
intervals.