5 calibration information, 1 calibration sheet, 2 poles and zeroes – Guralp Systems CMG-5T User Manual
Page 21

Operator's Guide
5 Calibration information
Every CMG-5T is supplied with a comprehensive calibration pack
detailing the characteristics of the sensor.
5.1 Calibration sheet
The calibration sheet provides the measured acceleration
responsivities over the flat portion of the sensor frequency response, in
units of volts per metre per second squared (V/ms
-2
). Because the
sensor produces outputs in differential form (also known as push-pull
or balanced output), the signal received from the instrument by a
recording system with a differential input will be twice the true value.
For example, the calibration sheet may give the acceleration
responsivity as “2 x 0.50 V/ms
-2
”, indicating that this factor of 2 was
not included in the value given.
Caution: You must never ground any of the differential outputs. If you
are connecting to a single-input recording system, you should use the
signal ground line as the return line.
5.2 Poles and zeroes
The poles and zeroes table describes the frequency response of the
sensor. If required, you can use the poles and zeroes to derive the true
ground motion mathematically from the signal received at the sensor.
The 5T is designed to provide a flat response (to within 3dB) over its
passband. For example, the following curve describes the frequency
response of a 100 Hz sensor:
December 2009
21