beautypg.com

3 calibration, 1 absolute calibration, 2 relative calibration – Guralp Systems CMG-5T User Manual

Page 12

background image

CMG-5T

3 Calibration

The 5T is supplied with a comprehensive calibration document, and it

should not normally be necessary to calibrate it yourself. However,
you may need to check that the response and output signal levels of

the sensor are consistent with the values given in the calibration
document.

3.1 Absolute calibration

The sensor's response (in V/ms

-2

) is measured at the production stage

by tilting the sensor through 90 ° and measuring the acceleration due

to gravity. Local g at the Güralp Systems production facility is known
to an accuracy of 5 digits. In addition, sensors are subjected to the

“wagon wheel” test, where they are slowly rotated about a vertical
axis.

The response of the sensor traces out a sinusoid over time, which is

calibrated at the factory to range smoothly from 1g to –1g without
clipping.

3.2 Relative calibration

The response of the sensor, together with several other variables, is
measured at the factory. The values obtained are documented on the

sensor's calibration sheet. Using these, you can convert directly from
voltage (or counts as measured in Scream!) to acceleration values and

back. You can check any of these values by performing calibration
experiments.

Güralp sensors and digitizers are calibrated as described below.

12

Issue E