Guralp Systems CMG-5T User Manual
Page 14

CMG-5T
2. Select the calibration channel corresponding to the
instrument, make any other choices you require, and click
Inject now. A new data stream, ending Cn (n = 0 – 7) or MB,
should appear in Scream!'s main window containing the
returned calibration signal.
3. Open a Waveview window on the calibration signal and the
returned streams by selecting them and double-clicking. The
streams should display the calibration signal combined with
the sensors' own measurements. If you cannot see the
calibration signal, zoom into the Waveview using the scaling
icons at the top left of the window or the cursor keys.
4. If you need to scale one, but not another, of the traces, right-
click on the trace and select Scale.... You can then type in a
suitable scale factor for that trace.
5. Click on Ampl Cursors in the top right hand corner of the
window. A white square will appear inside the Waveview at
the top left. This is in fact two superimposed cursors.
6. Drag one cursor down to be level with the lowest point of the
signal trace.
7. Drag the other down to be level with the highest point. In the
following example, a step function of 1 minute duration has
been applied to the Z3 stream. Note that ground movements
continue to be observed, superimposed on the returning
calibration signal.
14
Issue E