GBS Elektronik MCA166-USB User Manual
Page 69

MCA Data Setup Commands:
Command name
CMD_SET_TDF
Description
The command sets the TDF parameter.
tdf:
TDF(100 ... 3000, default 800)
Command syntax
Format
integer
integer
integer
long
integer
Parameter
preamble
command
tdf
0
end flag
Byte String (HEX)
A5
5A
61
00
tdf
l
tdf
h
0
0
0
0
B9
9B
Used setup value
None
Affected setup value
tdf
MCA state
STATE_READY
Command name
CMD_SET_UF6_ROI
Description
The command sets the begin and the end of a ROI used by the other UF6
commands.
r:
ROI number
b:
LLD<= b < e
e:
b < e <= ULD, (e-b) < 500
Command syntax
Format
integer
integer
integer
integer
integer
integer
Parameter
Preamble
command
r
b
e
end flag
Byte String (HEX)
A5
5A
64
00
r
00
b
l
b
h
e
l
e
h
B9
9B
Used setup value
LLD, ULD
Affected setup value
roi_begin[n], roi_end[n]
MCA state
STATE_READY
Command name
CMD_SET_USER_DATA
Description
The command stores a 32 bit value(f) in the MCA parameter memory(64
entries).
e: 0 ... 63
parameter number
p: 0 ...
long or float parameter
Command syntax
Format
integer
integer
integer
long or float
integer
Parameter
preamble
command
e
P
end flag
Byte String (HEX)
A5
5A
57
00
e
00
p
l
p
l
p
h
p
h
B9
9B
Used setup value
None
Affected setup value
None
MCA state
STATE_READY