GBS Elektronik MCA166-USB User Manual
Page 46

MCA166-USB
MCA Stabilisation Commands:
Name
MMCA_SET_STABILISATION
Description
The command sets the peak stabilisation.
1. Parameter
Stabilisation flag
Integer
value
0 turns stabilisation off
1 turns stabilisation on (keep peak at current
position)
>1 stabilisation channel (move peak to the
stabilisation
channel and hold peak at the
stabilisation channel)
2. Parameter
Peak ROI begin
Integer
value
LLD<= begin < end
3. Parameter
Peak ROI end
Integer
value
begin < end <= ULD, (end-begin) < 250
Results
-
Return Code
see MMCA_RESET command
Name
MMCA_SET_STAB_PARAM
Description
The command sets the peak stabilisation parameters.
1. Parameter
Stabilisation time
Integer
value
default 10 sec interval
2. Parameter
Stabilisation area
long value
default 25000
Results
-
Return Code
see MMCA_RESET command
MCA Detector Power Commands:
Name
MMCA_SET_PREAMPLIFIER_POWER
Description
The command sets the preamplifier power.
1. Parameter
Preamplifier switch integer value
0X80 -24V
0X40 +24V
0X20 -12V
0X10 +12V
Results
-
Return Code
see MMCA_RESET command
Name
MMCA_SET_HIGH_VOLTAGES
Description
The command sets the detector high voltage and controls the HV-inhibit-signal.
1. Parameter
High voltage
integer value
0 ... 3000
2. Parameter
High voltage inhibit long value
i = 0 Inhibit off
i = 1 ”Canberra-/ DGF-mode”, HV shut
down
if inhibit signal high
i =-1 ”Ortec-mode”, HV shut down
if inhibit signal low
Results
-
20 The high voltages will ramp up or down by the MMCA.
46