13 miscellaneous – GBS Elektronik MCA166-USB User Manual
Page 14

MCA166-USB
3.13 Miscellaneous
For communication and evaluation some libraries for MS Windows and Linux are
available. This may help the make own application programs or to integrate the
MCA166-USB in an existing system.
4 MCA166-USB parameters for use with different detectors
This table shall just give an overview of possible settings useful with the MCA166-USB. It
is far from being complete. For actual settings with your detector, see the detectors
manual.
Table 2: Parameters for different detectors
T
y
p
e
M
o
d
e
l
M
a
n
u
fa
c
t.
H
V
(
V
)
+
1
2
V
-1
2V
+
2
4
V
-
24
V
In
p
u
t
s
ig
n
a
l
p
o
la
ri
ty
C
h
a
n
n
e
ls
G
a
in
f
o
r
2
M
e
V
f
u
ll
ra
n
g
e
S
h
a
p
in
g
ti
m
e
(
µ
s
)
CdZnTe SDP/Z/60
Ritec
+300
x
x -
-
neg
1024 90
1
CdZnTe SDP/Z/20
Ritec
+500
x
x -
-
neg
1024 80
1
CdZnTe CZT500
Ritec
+600
x
x -
-
pos
1024 128
1
NaI
Scintiblock
Crismatec +500
pos
512
7.2
1
NaI
25825
Scionix
+550
x
x -
-
pos
512
90
1
NaI
40*40
Amcrys-h -800
x
x
pos
512
14
1
CsI
Sc1010
Eurorad
-
x
x -
-
neg
256
160
1
HPGe
GMX-20190-S Ortec
-3000
x
x x
x
neg
4096 18
2
HPGe
GEM 15-190
Ortec
+3000 x
x x
x
pos
4096 24
2
HPGe
GR 2018
Canberra -3000
neg
4096 11
2
HPGe
GL 0310
Canberra -2000
x
x x
-
neg
4096 2.3
2
Si
planar
7865S
Ortec
-1000
neg
4096 1.9
2
HPGe
EGC 30-190R Eurisys
-3000
pos
4096 52
2
HPGe
EGPC 25-185 Eurisys
+2000
neg
4096 35
2
14