GBS Elektronik MCA166-USB User Manual
Page 41
Name
MMCA_SET_ADC_RES_DISCR
Description
The command sets the ADC resolution and the software discriminator range.
1. Parameter
Channels
integer value
128, 256, 512, 1024, 2048 or 4096
2. Parameter
LLD
integer value
0 <= LLD < ULD
3. Parameter
ULD
integer value
LLD < ULD <(Channels-(Channels>>5))
Results
-
Return Code
see MMCA_RESET command
Name
MMCA_SET_ROI
Description
The command sets the begin and end of the ROI for the preset integral and area.
1. Parameter
ROI begin channel
Integer value
LLD <= begin < end
2. Parameter
ROI end channel
Integer value
LLD < end <= ULD
Results
-
Return Code
see MMCA_RESET command
Name
MMCA_SET_REPEAT
Description
The command sets the number of sweeps for repetitive measurement.
1. Parameter
Sweeps
Integer value
0 ... 65535
Results
-
Return Code
see MMCA_RESET command
Name
MMCA_SET_MCS_CHANNEL
Description
The command sets the number of channels for the MCS and Rate mode.
1. Parameter
MCS channels
Integer value
0 (only for Rate mode), 128, ... 4096
Results
-
Return Code
see MMCA_RESET command
Name
MMCA_SET_TIME_PER_CHANNEL
Description
The command sets the dwell time per channel for the MCS mode.
1. Parameter
Time
Integer value
1 ... 65535
Results
-
Return Code
see MMCA_RESET command
14 It is the real time per channel multiplied with 10 msec.
41